YSI 3100-230, Conductivity Instrument, Lab, 230V

SKU:
YSI 031037
$935.75
Shipping:
Calculated at Checkout

Description

The YSI 3100 provides accuracy and reliability in measuring conductivity, salinity and temperature.

The instrument uses forced-current measurement technology, and other features including direct-reading digital display, adjustable temperature coefficient, reference temperature, and automatic temperature compensation.

The YSI 3100 uses a line of NIST-traceable conductivity cells with built-in temperature sensors, assuring the most accurate measurements.

Ensure even more accurate readings by using YSI conductivity solutions and the 3166 Resistor set as part of your QA/QC procedures.

Features:

  • Provides high-accuracy measurements for basic conductivity needs
  • Uses forced-current technology
  • Measures conductivity, salinity and temperature
  • Easy-to-read LCD display
  • CE compliant

Specifications:

YSI 3100 Instrument General Specifications

Calibration single point
Cell Connector 7-pin Mini Din
Cell constant, cm-1 0.01, 0.1, 1, 10 cm-1
Certifications UL, CSA, CE
Dimensions Size 9 by 9.5 by 4.4 in (22.9 by 24.1 by 11.2 cm)
Weight 2.6 lbs (1.1 kg)
Display LCD
Operating Environment 95% relative humidity, non-condensing
Platinizing Capability Included
Power AC, 115V, 220V
Temperature Compensation
Method Linear
Reference temperature
15 - 25°C
Temperature Coefficient 0 - 4%/°C
Water Resistance
The Model 5000 and 5100 are designed exclusively for indoor use and are NOT waterproof.
Warranty
2-year instrument

 

YSI 3100 Specifications (Instrument Only)

Conductance
Range Accuracy Resolution Frequency
0 to 49.99 µS/cm1 ±0.50% full scale 0.01 µS/cm 70 Hz
0 to 499.9 µS/cm
±0.50% full scale 0.1 µS/cm 70 Hz
0 to 4999 µS/cm2 ±0.50% full scale 1 µS/cm 240 Hz
0 to 49.99 mS/cm3
±0.50% full scale 0.01 mS/cm 1562 Hz
0 to 499.9 mS/cm4
±0.50% full scale 0.1 mS/cm
1562 Hz
1 Requires a cell constant of K=0.01, K=0.1 or K=1.
2 Requires a cell constant of K=0.1, K=1 or K=10.
3 Requires a cell constant of K=1 or K=10.
4 Requires a cell constant of K=10.

 


Range Accuracy Resolution
Salinity 0-80 ppt (NaCl) 2% or ±0.1 ppt 0.1 ppt
Temperature
-5 to +95°C
±0.1°C + 1 lsd 0.1°C

 


Conductivity Cell General Measurement Ranges

 

Conductivity Cell General Specifications

3200 Series Cells with built-in temperature sensors


Model cgs Cell Type
S.I. Cell Constant
Cell Constant Material Overall Length (mm) Max O.D. (mm) Chamber I.D. (mm) Chamber Depth (mm) Volume
A 3252 dip 1.0/cm 100/m ABS plastic 146
13
10
20
 
B 3253 dip, micro
1.0/cm 100/m Pyrex 7740 178
13
10
51
 
C
3254 fill 1.0/cm 100/m Pyrex 7740 135
19
11
83
5 mL
D 3255 flow 0.1/cm 10/m Pyrex 7740 146
25
21
76
30 mL
E 3256 dip 0.1/cm 10/m Pyrex 7740 159
25
21
52
 


3400 Series Cells without built-in temeprature sensors*


  Model cgs Cell Type S.I. Cell Constant Cell Constant Material Overall Length (mm)
Max O.D. (mm)
Chamber I.D. (mm) Chamber Depth (mm) Volume
F
3401
dip 1.0/cm 100/m Pyrex 7740 191
25 21 76  
G
3402
dip 0.1/cm 10/m Pyrex 7740 159
25 21 52  
H
3403
dip 1.0/cm 100/m Pyrex 7740 178 13 10 51  
I
3417
dip 1.0/cm 100/m ABS plastic 146 13 10 20  
I 3418 dip 0.1/cm 10/m ABS plastic 159 13 10 30  
J
3440
dip 10/cm 1000/m Pyrex 7740 203 13 2 86  
*Requires a YSI 3232 Cell Adaptor for use with YSI 3100 and 3200 Conductivity Instruments. For automatic temperature compensation, use a YSI 3220 on the 3200

Extra Information

Brand:
YSI
Weight:
5.00 LBS