Product Description

2020ComboPRO Intrinsically Safe Hand-Held Photoionization Detector with VOC Field Kit 

Includes 10.6eV high performance lamp, 115V AC adapter, and VOC Field Kit (carrying case (MX700260), spare 8-hour rechargeable battery (MX700010), calibration gas regulator (MX350006), ProComm software CD (MX790116), and printer/computer cable (MX750120). Note: Field Kit does not include calibration gas. 

Basic accessories for all 2020PRO series PID’s include specified UV lamp and AC charger (115 or 220 VAC), plus 8 hour rechargeable battery pack, 6.7 in. (17cm) Teflon sample probe with spring relief, 10 Fluoropore membrane filters, adjustable wrist strap, multi tool, user reference card, and user manua (CD). Classified Intrinsically Safe for use in Class I, Division 1, Groups A, B, C, and D hazardous locations and ATEX II 2 G EEx ib IIC T4.

2020ComboPRO is Intrinsically Safe to North American and ATEX standards for use in potentially explosive environments.Expanded internal data logging allows to record up to 200 hours of 1 minute interval sampling points. Zero calibration through the use of an activated charcoal pre-filter tube developed for Photovac by Dräger Safety AG.

  • Fenceline and perimeter monitoring
  • Confined space pre-entry
  • Solvent storage and piping
  • Transportation vessels
  • Storage tank maintenance
  • Site characterization
  • Ambient air monitoring
  • Emergency response and HazMat
  • Petroleum product(s) tank entry
  • Soil headspace screening
  • Health and safety monitoring
  • Surveying for buried chemical waste
  • Aircraft fuel tank entry
  • Meth Lab Cleanup


Instant on photoionization detector with standard 10.6 eV lamp, and optional 11.7 eV lamp.

9” (228.6 mm) long x 3” (76.2 mm) deep x 4.25” (107.9 mm) wide at display tapering to 2.6” (66.6 mm) at handle

1.9 pounds (0.86 kg)

Two dedicated keys (ON/OFF and CAL) and three menu keys

Large 128 x 64 graphic LCD, 6-line

Datalogger Memory
200 hours at 1 minute interval sampling

Serial Output
RS-232, 9600 baud, 8 data bits, no parity for connection to Windows-based PC

Audio Output
85 dB on alarm

Inlet Connection
1/8"(3.175 mm) compression fitting

Operating Temperature Range
Operating temperature: 0°C to 50°C (32°F to 122°F)

I/S Certified temperature: 0°C to 40°C (32°F to 104°F)

Operating Humidity Range
0 to 95% relative humidity (non-condensing)

Operating Humidity Range with Optional Humidity Pre-filter Tube
0 to 100% relative humidity (non-condensing)

Operating Concentration Range
VOC Mode:
0.1 ppm to 100 ppm (isobutylene equivalent), 100ppm to 10,000ppm (isobutylene equivalent)

Benzene Mode: 0.1 ppm to 40 ppm (isobutylene equivalent)

Response Time
Less than 3 seconds, to 90%

Low Detection Limit
0.1ppm Isobutylene

Battery Capacity
8 hours

Intrinsic Safety
Class 1, Division 1, Groups A, B, C, & D

ATEX II 2 G EEx ib IIC T4 (TA = 0°C to +40°C)

Class B digital device, pursuant to Subpart B

Class B of Part 15 of the FCC rules

English (standard).  One of the following can be added as an option: Spanish, German, French, or Italian